Scigiene's Wireless Gen 2 High Temperature Sensor uses a PT1000 RTD to accurately measure temperatures from -50°C to +370°C. It is perfect for high temperature critical applications such as ovens, heaters, furnaces and boilers. User customization allows you to set the frequency of sensor readings and create notifications from the system when certain criteria have been met or exceeded. Scigiene's wireless sensors are ideal for pharmaceutical, medical and other high tech applications where loss of data is unacceptable. These also have an amazing range of over 1200 ft. with 10 year battery life and can be remotely upgraded via network connections. These are truly the most advanced wireless sensors available.
Features:
* Actual range may vary depending on environment.
**Battery life is determined by sensor reporting frequency and other variables. Other power options are also available
Specifications:
Temperature
Measurement |
Range- Leaded
sensor |
-50°C to
370°C (-58°F to 700°F) |
Accuracy @ 0°C
(32°F) |
± 3.3°C
(±5.9°F) | |
Calibrated
accuracy @ 0°C (32 °F) |
± 0.5°C (±
0.9°F) | |
Resolution |
0.1°C
(0.18°F) | |
Response time |
50 seconds
(10-second time constant)1 | |
Leaded
Probe |
Transducer
type |
Glass-coated
platinum RTD, IEC 751-95 Class A |
|
Tip
dimensions |
4.76 mm
(0.1875") diameter by 35 mm (1.38") |
|
Tip material |
Type 304
stainless steel |
|
Cable
material |
Glass Braid |
|
Cable
diameter |
1.5 x 2.4 mm
(.060" x 0.095") |
|
Cable length |
Standard:
0.9m (3 ft) Option: 3m
(10ft), 7.5m (25 ft) |
Gen 2 Wireless |
Data logging |
Sensor logs
2000 to 4000 readings if gateway connection is lost (non-volatile flash,
persists through power cycling): 10-minute Heartbeats = ~22 days - 2-hour
Heartbeats = ~266 days |
Wireless
protocol |
Gen 2 Proprietary Frequency-Hopping Spread Spectrum (FHSS) | |
Wireless transmission
power (EIRP) |
50 mW
(900MHz), 25 mW (868 MHz), 10 mW (433 MHz) | |
Wireless
range |
2,000+ ft.
through 18+ walls with the Gen 2 XL® Gateway | |
Security |
Encrypt-RF ®
(256-bit key exchange and AES-128 CTR) | |
General |
Battery
voltage range |
2.0 to 3.8
VDC |
Operating
altitude (non-pressurized environments) |
-15.2 to
1,982 m (-50 to 6,500 ft)2 | |
Storage
altitude (non-pressurized environments) |
-15.2 to
3,048 m (-50 to 10,000 ft)2 | |
Operating
humidity |
5 to 85% RH
(non-condensing) | |
Certifications |
900 MHz
sensors: FCC ID: ZTL-G2SC1 and IC: 9794A-G2SC1. |
|
1. Response time defined as five time constants for
99.3% of actual temperature. 2. Operating and storage altitude without DC power
supply is -30.48 to 9144 m (-100 to 30000 ft). |