The SciTemp140X2
series of dual probe high temperature data loggers are comprised of a stainless
steel data logger body and offers multiple combinations of a 2 inch rigid
transitional diameter probe to measure ambient temperature, a flexible probe or
a bendable stainless steel probe.
The FP
models offer lightweight, flexible RTD probes coated with PFA insulation and
come in 6 inch, 12 inch, 36 inch and 72 inch lengths and can be combined with a
bendable probe or rigid probe. The PT models combine a flexible probe or rigid
probe with a 24 inch bendable probe made of stainless steel with either a 1
inch or 5 inch probe sheath at the tip. The TD models combine the 2 inch rigid
transitional diameter probe with either the flexible probe or bendable probe.
Using
the latest SciTemp Software, starting, stopping and downloading the SciTemp140X2
data is simple and easy. To use, simply place the SciTemp140X2 in the SD900319-00
or SD900325-00 docking station (sold separately). Using the software, an
immediate or delay start can be chosen, as well as the reading interval. Start
the data logger, remove it from the docking station and the device is ready to
be deployed. Graphical, tabular and summary data is provided for analysis and
data can be viewed in °C, °F, K or °R. The data can also be automatically
exported to Excel® for further calculations.
Features:
Specifications:
Temperature Specifications |
|
Temperature
Sensor |
SciTemp140X2-
FPST:
Two Flexible RTD Probes SciTemp140X2-
FPST-PT:
Flexible RTD Probe and Bendable RTD Probe SciTemp140X2-TD-PT: Rigid RTD Probe
with a Bendable RTD Probe SciTemp140X2-TD-FPST: Rigid RTD Probe
with a Flexible RTD Probe |
Probe Measurement
Range |
Rigid Probe: -200 °C to +260
°C (-328 °F to +500 °F) Bendable Probe: -200 °C to +350
°C (-328 °F to +662 °F) Flexible Probe: -200 °C to +260
°C (-328 °F to +500 °F) *BODY OF LOGGER CANNOT
EXCEED 140 °C |
Temperature Resolution |
0.01 °C (0.02 °F) |
Calibrated
Accuracy |
±0.1 °C (+20 °C
to +140 °C); 0.18 °F (+68 °F to +284 °F) ±0.3 °C (-20 °C
to 19.99 °C); 0.54 °F (-4 °F to +67.98 °F) ±0.4 °C (-40 °C
to -20.01 °C); 0.72 °F (-40 °F to +68.018 °F) |
General Specifications |
|
Start Modes |
Software
programmable immediate start Delay starts up
to 18 months in advance |
Stop Modes |
Manual or Timed
(specific date and time) |
Trigger Settings |
High and Low
limits may be set. Once data meets or exceed sets limits, the device will
record to memory. Bi-level start and stop triggers can also be programmed.
Users can specify the number of readings to take after the device triggers.
(Triggering on channel #1 only) |
Memory |
16,128 readings |
Readings in
Trigger Settings Mode |
5,376 readings |
Real Time
Recording |
May be used with
PC to monitor and record data in real time |
Password
Protection |
An optional
password may be programmed into the device to restrict access to
configuration options. Data may be read out without the password |
Wrap Around |
Yes |
Reading Rate |
1 reading every
second up to 1 reading every 24 hours |
Calibration |
Digital
calibration through software |
Calibration Date |
Automatically
recorded within device |
Battery Type |
3.6 V
high-temperature lithium battery included; user replaceable |
Battery Life |
1 year typical (1
minute reading rate at 25 °C/ 77 °F) |
Data Format |
Date and time
stamped °C, K, °F or °R |
Time Accuracy |
±1 minute/month
at 25 °C (77 °F) Extended
Operation: ±20 minutes/month at 140 °C (±450 ppm) |
Computer
Interface |
SD900319-00 OR SD900325-00
USB docking station required; 125,000 baud |
Operating System
Compatibility |
Windows XP SP3 or
later |
Software
Compatibility |
SciTemp Standard
Software version 4.2.1.0 or later SciTemp Secure
Software version 4.2.0.0 or later |
Operating
Environment |
-40 °C to +140 °C
(-40 °F to +284 °F) 0 %RH to 100 %RH,
0.002 PSIA to 100 PSIA |
IP Rating |
IP68 |
Dimensions (Body) |
1.89 in x 0.97 in
x 0.97 in |
Dimensions
(Probe) |
6 inch Flexible
Probe:
6 in x 0.13 in 12 inch Flexible
Probe:
12 in x 0.13 in 36 inch Flexible
Probe:
36 in x 0.13 in 72 inch Flexible
Probe:
72 in x 0.13 in |
1 Inch PT
Bendable Probe: Probe tip: 1.7 in x 0.125
in dia. Bendable portion: 22 in x 0.062 in
dia. |
|
5 Inch PT
Bendable Probe: Probe tip: 4.8 in x 0.125
in dia. with 1 in x 0.188 in dia. handle Bendable portion: 22 in x 0.062 in
dia. |
|
TD Rigid Probe: 2.0 in x 0.125 in
dia. (0.188 in transitional dia.) 51 mm x 3.2 mm
dia. (4.8 mm transitional dia.) |
|
Weight |
SciTemp140X2-FPST:
4.1 oz (115 g) SciTemp140X2-FPST-PT:
3.9 oz (110 g) SciTemp140X2-TD-PT:
3.0 oz (85 g) SciTemp140X2-TD-FPST:
3.5 oz (100 g) |
Material |
Body: 316
Stainless Steel, Radel Bendable Probe:
316 Stainless Steel Flexible Probe:
PFA Insulated Cable with Stainless Steel Tip |
Approvals |
CE |
Model & Accessory Options:
Part # |
Description |
SD902335-00 |
SciTemp140X2-FP-6
- High
Temperature Dual Probe Data Logger with two 6" Flexible RTD Probes
with Steel Tip |
SD902334-00 |
SciTemp140X2-FP-6-FP-12
- High
Temperature Dual Probe Data Logger with one 6" Flexible RTD Probe
with Steel Tip and one 12" Flexible RTD Probe with Steel Tip |
SD902328-00 |
SciTemp140X2-FP-6-FP-36
- High
Temperature Dual Probe Data Logger with one 6" Flexible RTD Probe
with Steel Tip and one 36" Flexible RTD Probe with Steel Tip |
SD902331-00 |
SciTemp140X2-FP-12
- High
Temperature Dual Probe Data Logger with two 12" Flexible RTD Probes
with Steel Tip |
SD902332-00 |
SciTemp140X2-FP-12-PT-1
- High
Temperature Dual Probe Data Logger with one 12" Flexible RTD Probe
with Steel Tip and one 24" Bendable RTD Probe with a 1" probe
sheath |
SD902333-00 |
SciTemp140X2-FP-24
- High
Temperature Dual Probe Data Logger with two 24" Flexible RTD Probes
with Steel Tip |
SD900265-00 |
SciTemp140X2-FP-36
- High
Temperature Dual Probe Data Logger with two 36" Flexible RTD Probes
|
SD902092-00 |
SciTemp140X2-FP-36-KR
- High
Temperature Dual Probe Data Logger with two 36" Flexible RTD Probes
and Key Ring End Cap |
SD900267-00 |
SciTemp140X2-FP-36-PT-1
- High
Temperature Dual Probe Data Logger with one 36" Flexible RTD Probe
and one 24" Bendable RTD Probe with a 1" probe sheath |
SD900271-00 |
SciTemp140X2-FP-36-PT-5
- High
Temperature Dual Probe Data Logger with one 36" Flexible RTD Probe
and one 24" Bendable RTD Probe with a 5" probe sheath |
SD900271-00 |
SciTemp140X2-FP-72
- High
Temperature Dual Probe Data Logger with two 72" Flexible RTD Probes |
SD900273-00 |
SciTemp140X2-FP-72-PT-1
- High
Temperature Dual Probe Data Logger with one 72" Flexible RTD Probe
and one 24" Bendable RTD Probe with a 1" probe sheath |
SD9002753-00 |
SciTemp140X2-FP-72-PT-5
- High
Temperature Dual Probe Data Logger with one 72" Flexible RTD Probe
and one 24" Bendable RTD Probe with a 5" probe sheath |
SD900281-00 |
SciTemp140X2-TD-PT-1
- High
Temperature Dual Probe Data Logger with one 2" TD Probe and one
24" Bendable RTD Probe with a 1" probe sheath |
SD900283-00 |
SciTemp140X2-TD-PT-5
- High
Temperature Dual Probe Data Logger with one 2" TD Probe and one
24" Bendable RTD Probe with a 5" probe sheath |
Accessories: |
|
SD900319-00 |
Docking station
with USB cable |
SD900325-00 |
6 Port, Multiplexer
docking station with USB cable |
SD900097-00 |
Replacement
battery for the SciTemp140-FR |