This is a simple, fast and affordable mini titrator designed for testing total acidity levels in wine. Based on an acid-base titration method, this mini titrator uses a preprogrammed method of analysis with a powerful algorithm that determines the completion of the titration reaction by use of a specialized wine pH electrode. This mini titrator is supplied complete with all the materials necessary to perform low and high range measurements of total acidity. All chemicals are premixed and prepackaged including standardized titrants, reagents and pump calibration solution. There is no need for volumetric glassware or analytical balances.
Application-Specific pH Electrode
The HA-HI84502 is supplied with a glass body pH electrode featuring CPS™technology to prevent the clogging of the reference junction. Conventional electrodes may clog quickly in biological samples that have a high solids content such as wine must. By design, the pH electrode utilizes a polytetrafluoroethylene (PTFE) sleeve junction which controls a steady, predictable flow of electrolyte solution, keeping the junction open. The hydrophobic properties of PTFE repel wetness and coatings.
Piston Driven Dosing Pump
The heart of the HA-HI84502 is the piston driven burette pump. This type of dosing system uses a motor in which each dose is very precisely controlled and the volume dispensed is accurately determined. The piston driven burette is controlled dynamically so that the volume of titrant being dosed is automatically adjusted based on a voltage response of the previous dose. This type of dosing speeds up the titration process by allowing for more titrant to be dosed at the start of the titration and then very small doses as the endpoint is reached.
Automatic Stirrer
The built-in stirrer is automatically maintained at a speed of 600 rpm, regardless of the viscosity of the solution being titrated.
Easy-to-Use Interface The HA-HI84502 has an intuitive user design with clearly defined keys and a large screen that is easy to navigate. The meter has a built-in tutorial mode that, when enabled, will walk the user step by step through the titration process. A dedicated HELP key is always available to allow access to content-specific information during both calibration and titration.
Range | Low Range: 0.1 to 5.0 g/L of tartaric acid; High Range: 4.0 to 25.0 g/L of tartaric acid |
Resolution | 0.1 g/L (ppt) |
Accuracy (@25ºC/77ºF) | +/- 3% of reading or +/-0.1 g/L, whichever is greater |
Sample Volume | Low Range: 10 mL; High Range: 2 mL |
Methods | acid base titration |
Principle | endpoint titration: 7.00 pH or 8.20 pH |
Pump Speed | 10 mL/min |
Stirring Speed | 600 rpm |
pH Range | -2.0 to 16.0 pH; -2.00 to 16.00 pH |
pH Resolution | 0.1 pH / 0.01 pH |
pH Accuracy (@25ºC/77ºF) | +/-0.01 pH |
pH Calibration | one, two, or three-point calibration, four available buffers (4.01, 7.01, 8.20, 10.01) |
Temperature Compensation | manual or automatic |
mV Range | -2000.0 to 2000.0 mV |
mV Resolution | 0.1 mV |
mV Accuracy (@25ºC/77ºF) | +/-1.0 mV |
Temperature Range | -20.0 to 120.0°C; -4.0 to 248.0°F; 253.2 to 393.2 K |
Temperature Resolution | 0.1°C; 0.1°F; 0.1K |
Temperature Accuracy (@25ºC/77ºF) | +/-0.4°C; +/-0.8°F; +/-0.4 K |
Logging Data | up to 400 ( 200 titrations, 200 pH/mV readings) |
Electrode | HA-HI1048B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included) |
Temperature Probe | HA-HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
Power Supply | 12 VDC adapter (included) |
Ordering Information | HA-HI84502 is supplied with HA-HI1048B pH electrode, HA-HI7662-T temperature probe, HA-HI7082 electrode fill solution (30 mL), HA-HI84502-50 (230 mL) titration solution, HA-HI84502-55 pump calibration standard (120 mL), 100 mL beakers (2), dosing pump valve, 2000 μL automatic pipette with plastic tips (2), 5 mL syringe, 1 mL plastic pipette, tube set (aspiration tube with titrant bottle cap and dispensing tube with tip), stir bar, cleaning solution sachets for wine deposits, cleaning solution |
Part # | Description |
HA-HI1048B | pH Electrode with Clogging Prevention System (CPS™) and BNC Connector |
HA-HI7662-T | Replacement Stainless Steel Temperature Probe |
HA-HI84502-50 | Titrant for Titratable Acidity in Wine Mini Titrator, 230 ml bottle |
HA-HI84502-55 | Pump Calibration Standard for Titratable Acidity in Wine Mini Titrator, 120 ml bottle |
HA-HI70300L | Electrode Storage Solution (500 mL) |
HA-HI7082L | Electrolyte Solution 3.5M KCl (460 mL) |
HA-HI7004L | pH 4.01 Calibration Solution (500 mL) |
HA-HI7007L | pH 7.01 Calibration Solution (500 mL) |
HA-HI70082M | pH 8.20 Calibration Buffer Solution (230 mL) |
HA-HI7010L | pH 10.01 Calibration Solution (500 mL) |
HA-HI70635L | Cleaning Solution for Wine Deposits (500 mL) |
HA-HI700635P | Cleaning Solution for Wine Deposits (25 x 20 mL Sachets) |
HA-HI700636P | Cleaning Solution for Wine Stains (25 x 20 mL Sachets) |