The 3C\TEMP-USB -200C is a reliable and cost-effective digital temperature recorder for all industrial and commercial applications. It has one sensor for ambient temperatures inside the case, and one external RTD stainless-steel probe that monitors temperatures from -200°C (-328°F) up to +72°C (+161.6°F). It is ideal for monitoring all of your cryogenically stored stem cells and biologic materials. These high-precision laboratory-grade products will operate up to 1.5 years without a battery change. The LCD display provides real time viewing of ambient and probe temperatures prior to downloading to a computer.
Features:
Specifications:
Memory/Storage | 8k memory (approximately 8,000 measurements) |
External Sensor/Probe count and connector type | 1, Soldered Cryogenic Sensor (external probe): RTD Temperature Range: -200°C to +300°C / -328°F to +572°F Temperature Accuracy: ±0.5°C at 0° to 30°C, ±2.0°C at extremes / ±0.9°F at 32° to 86°F, ±3.6°F at extremes Resolution: 0.1°C / 0.18°F |
Internal Sensor | Ambient Temperature sensor (inside case): Wired chip in glass thermistor Temperature Range: -29°C to +72°C / -20°F to +161°F Temperature Accuracy: ±0.2°C at 0° to 10°C, ±1.0°C at extremes / ±0.4°F at 32° to 50°F, ±1.8°F at extremes Resolution: 0.1°C / 0.18°F |
Alarm Programming | User configurable. High and low alarm limits can be set by temperature |
Alarm Modes | Exceed Limit. See Alarms for more information |
Alarm Reset Options | Alarm is reset when logger is erased and reprogrammed |
Alarm Notifications | On screen text. See Alarms for more information. |
Audible Alarm | None |
Display | LCD screen 9.45cm(l) x 7.9cm(w) / 3.72in(l) x 3.1in(w) |
Indicators: | On screen text |
CE Certified: | Yes |
Water-resistant | IP66 rated |
Internal Firmware, Internal Clock Accuracy | Internal Clock, Start/Stop time, Interval, User description, Alarm values, Recording status, Firmware version, Serial number, Model. Clock Accuracy is better than 2 minutes per month on average at normal temperatures (-29°C to +72°C). For loggers directly exposed to temperatures at or below -80°C, the expected accuracy is ±15 minutes per month |
Measurement Intervals | Programmable from 2 seconds per reading. See Duration Chart for examples. |
Computer Interface (physical) | Mini-USB |
Wireless Connection and Range | None |
Remote Monitoring and Network Connection type | None |
Power Source, Battery Life and Replacement | One CR2450 Battery. 1 year operating life, and 3 year shelf life |