The SwabSURE Salmonella swab is a direct test for the
presumptive detection of Salmonella spp. from food
contact/preparation areas or manufacturing environments where regular
disinfecting/sanitizing is performed. Results are easily visualized using the
inbuilt colour change system giving clear results. Kit includes Pre-moistened
(in TSC Neutralizing Buffer) Breakpoint Swab for simple sample
transfer. ISO18593:2018 compliant when used a sample template.
Media inhibits/suppresses other background microorganisms
such as: Bacillus spp., Enterococcus spp., Micrococcus spp
and Gram negative organisms like Klebsiella spp, E.coli and
Pseudomonads. Neutralizing Swabs are also sold separately.
Ideal For: Detecting Salmonellae from surfaces
including food preparation areas, drains, floors etc. Routine Sampling for
Pathogens is a fundamental part of Good Manufacturing Practice
(GMP) and essential for adhering to a HACCP (Hazard Analysis Critical
Control Points) plan. This product can be used to compliment (or in conjunction
with) the ISO6579-1:2017 +A1:2020 testing/detection methodology.
2. SwabSURE is intended as a screening test. All positive
results should be confirmed by standard bacteriological or molecular methods.
3. It is recommended that after sampling food contact areas
are wiped to remove any residual traces of neutralizing solution. If used for
direct food contact the item must be removed/discarded after use (so not sold).
4. Has not been validated for uncleaned food surfaces. The
use of SwabSURE for these purposes may result in false positives or
spurious/unreadable results.
5. The absence of a colour change does not necessarily imply
the absence of target bacteria. Reasons for this may include:
a. Inadequate neutralization of sanitizing agents.
b. Numbers of target organisms are below the detection level
of the test.
c. The target organisms, although present may be severely
damaged and unable to be resuscitated.
6. A positive colour change may arise due to the
presence of excessively high numbers of competing organism species like Citrobacter spp.
producing a false positive result. The occurrence of false positive
results is an important indicator of suspect cleaning and sanitizing practices
due to concentration of background flora and should flag a review. Some
uncommon strains (e.g. S. arizonae) may be H2S negative
and could be missed though are usually associated with reptiles.
7. Long periods of Low/cold temperatures can cause
thickening/gel formation within the SS-L03 media. This characteristic is not
detrimental to the product's performance and once removed from cold
temperatures (brought to room temperature or incubated after sampling) the
media will return to liquid.
Shelf Life: 12 months - Components should not be
used after the expiry date printed on them or the carton label.
Storage: Store at 2-8°C when not in use. The
pre-moistened swabs should be stored at 4-25°C.
Sterilization Method: Terminally Sterilized
(detection media autoclaved, swab gamma irradiated)