The Kanamycin Test Kit is a single test kit, That’s rapid to detect antibiotic residues in milk and dairy products by using colloidal gold immunochromatography technology as little as 7-10 minutes for results.
Application
For raw milk, whole milk powder, Pasteurized milk.
Specifications
1. Specificity
Not cross-react with macrolides, trimethoprim, sulfonamides drugs, etc.
2. Limit of Detection
Refer to instruction.
Storage & Shelf-Life
At 2~8℃ out of sunlight, 12 months
Components (96T / box)
Instruction
1
Tube
12
Equipment’s (Option)
Timer
Incubator
Mini Pipette(200μL)
Tips for mini pipette
Reader
Interpretation
Visual inspection
1. If C line visible, interpret according to the instruction. T line compare
with C line to interpret positive or negative. (See picture above).
2. If C & T are invisible, it’s invalid detection, repeat testing
again. (See picture above).
Reader Interpretation
Refer to the instruction
Contents:
12 vials
(each containing 8 test strips and 8 micro wells) Total 96 tests.
Required
accessories:
Micropipettor
Micropipette
tips (volume ?)
Scale 0.1g
Solvent
(Ethanol /Methanol*)
Optional
accessories:
Vortexer
Microcentrifuge
The use of
these will allow you to speed up test time and/or increase accuracy.
*contact us
relative to product being tested to confirms correct solvent.
CTRL line is a quality indicator and
will always appear regardless of the T, B, C & S
Lines
results. If CTRL line
does not appear, please repeat the assay with a new strip.
Performance Information
Precautions and Warnings
1. Do
not mix sample tubes and devices from different lot numbers
2. Do
not use kits past their expiration dates.
3. Do
not touch the field of control/test line, before or while testing.
4. Keep
the kits in sealed when not in use.
5. The
kit is single-use, do not reuse.
Shelf Life: 12 months
Instructions for Use
Prepare
the milk sample, and the sample should be at room
temperature,
there is no need to heat or refrigerate.
1. Take
out the strip,
2. Drop
200ul into the microwell, repeatedly pipetting up
and
down 10 times, mix well with the reagent in the wells
completely.
3. Then
wait for 3 minutes
4.
Insert the test strip into the microwell until the botton of the strip touch
the bottom of the microwell.
5.
After 7 minutes, read the results.
If you
need to record the results then you must take the test strip, dry the
strip
and store with your records.